HiperS-320i Aberration-Free Imaging Spectrometer, 320 mm, On-Axis Turret
HiperS-320iDesign Tools
320 mm imaging spectrometer corrected for aberrations across the full focal plane, for multichannel spectroscopy.
HiperS-320i is a 320 mm imaging monochromator/spectrograph that applies full-focal-plane imaging correction to suppress astigmatism across the whole focal plane. An on-axis rotating grating turret and automatic focusing keep resolution and image quality consistent when gratings are changed, so intensity values are reliable over the full CCD. It is suited to LIBS, Raman, fluorescence and other molecular spectroscopy, including multichannel fiber-bundle measurements.
Key Features
- Full-focal-plane aberration correction for sharper peak shapes, more precise wavelengths and higher spatial resolution
- On-axis grating rotation increases grating utilization by 15% and keeps grating efficiency consistent
- Automatic focusing moves to the optimum focal plane whenever the grating is switched
- Built-in resolution enhancement module improves spectral resolution by up to about 20%
- Optional two entrance and two exit ports with automatic switching
- Multichannel imaging, demonstrated with a 9-core, 100 µm fiber bundle over a 3.8 mm image height
- Triple turret for up to three 68 × 68 mm gratings, from 150 to 2400 g/mm
Typical Applications
- Laser-induced breakdown spectroscopy (LIBS)
- Raman spectroscopy
- Fluorescence and photoluminescence
- Molecular spectroscopy
- Multichannel fiber-bundle spectroscopy
Model Overview
| Model | Entrance | Exit | UV Coating |
|---|---|---|---|
| HiperS-3204i | Side entrance slit | Dual CCD exit ports | – |
| HiperS-3205i | Side entrance slit | Front exit slit | – |
| HiperS-3206i | Side entrance slit | Front CCD exit port | – |
| HiperS-3207i | Side entrance slit | Dual exit slits | – |
| HiperS-3208i | Side entrance slit | Front CCD port + side exit slit | – |
| HiperS-3224i | Dual entrance slits | Dual CCD exit ports | – |
| HiperS-3227i | Dual entrance slits | Dual exit slits | – |
| HiperS-3228i | Dual entrance slits | Front CCD port + side exit slit | – |
| HiperS-3247i | Side entrance slit | Dual exit slits | Yes |
| HiperS-3248i | Side entrance slit | Front CCD port + side exit slit | Yes |
Related Products
- Omni-λ Series — Monochromators and Spectrographs
- Omni-iSpecT — VPH Raman Spectrograph
- Omni-λ300Di / Omni-λ500Di / Omni-λ180Di — Double monochromator / spectrograph
Browse all: Monochromators & Spectrographs · Spectroscopic Instruments
Models: HiperS-320i, HiperS-3204i, HiperS-3205i, HiperS-3206i, HiperS-3207i, HiperS-3208i, HiperS-3224i, HiperS-3227i, HiperS-3228i, HiperS-3247i, HiperS-3248i · Contact WaveQuanta for a quote, configuration advice and lead time.
| Focal Length | 320 mm |
| Aperture Ratio | F/4.4 |
| Scan Range (1200 g/mm) | 0–1430 nm |
| Resolution (1200 g/mm) | 0.06 nm (PMT, 10 µm slit); 0.12 nm (CCD, 15 µm pixel) |
| Dispersion (1200 g/mm) | 2.29 nm/mm |
| Wavelength Accuracy / Repeatability (1200 g/mm) | ±0.2 nm / ±0.025 nm |
| Step (1200 g/mm) | 0.005 nm |
| Stray Light | 5 × 10⁻⁵ |
| Focal Plane Size | 29 mm (W) × 14 mm (H) |
| Optical Axis Height | 156.3–171.3 mm |
| Grating Size | 68 × 68 mm |
| Grating Mount | On-axis triple turret |
| Slit Width | 0.01–3 mm continuous manual adjustment; motorized slit optional (not on front slit) |
| Slit Height | 2, 4 or 14 mm, switchable |
| Interface | USB 2.0 or RS-232 |
| Dimensions (L × W × H) | 591 × 460 × 260 mm |
| Weight | 30 kg |
| Supplied Configuration | Manual slits; gratings (up to 3), filter wheel and shutter selected separately |
| Dispersion by Grating (nm/mm) | 2400 g/mm: 0.91; 1800: 1.39; 1200: 2.29; 600: 4.87; 300: 9.97; 150: 20.13 |
| Scan Range by Grating (nm) | 2400: 0–715; 1800: 0–953; 1200: 0–1430; 600: 0–2860; 300: 0–4290; 150: 0–11440 |
| Slit Resolution by Grating (nm) | 2400: 0.03; 1800: 0.04; 1200: 0.06; 600: 0.12; 300: 0.24; 150: 0.48 |
| CCD Resolution by Grating (nm) | 2400: 0.048; 1800: 0.073; 1200: 0.12; 600: 0.256; 300: 0.523; 150: 1.06 |
| CCD Single-Shot Range, 30 mm CCD (nm) | 2400: 26.4; 1800: 40.3; 1200: 66.4; 600: 141.2; 300: 289.1; 150: 583.8 |
| Accuracy by Grating (nm) | 2400: ±0.1; 1800: ±0.15; 1200: ±0.2; 600: ±0.4; 300: ±0.8; 150: ±1.6 |
| Repeatability by Grating (nm) | 2400: 0.015; 1800: 0.02; 1200: 0.025; 600: 0.05; 300: 0.1; 150: 0.4 |
| Step by Grating (nm) | 2400: 0.0025; 1800: 0.0035; 1200: 0.005; 600: 0.01; 300: 0.02; 150: 0.04 |
| Test Conditions | Central wavelength 435.83 nm, 10 µm slit width; values typical |
Product Family: Monochromators & Spectrographs (5 variants)
| Product | SKU | Size | Clear Aperture | Operating Wavelength | |
|---|---|---|---|---|---|
| Omni-λ180Di, 300Di, 500Di Double Monochromators, Stray Light to 1×10⁻⁹ |
Omni-λ300Di / Omni-λ500Di / Omni-λ180Di
| — | — | — | Contact |
| Omni-iSpecT VPH Raman Spectrograph, F/2.3, 532 nm and 785 nm Versions |
Omni-iSpecT
| — | — | — | Contact |
| Triple Monochromator for Raman, Subtractive and Additive Dispersion Stages |
Triple Monochromator
| — | — | — | Contact |
| HiperS-320i Aberration-Free Imaging Spectrometer, 320 mm, On-Axis Turret Current |
HiperS-320i
| — | — | — | |
| Omni-λ Series Monochromators and Spectrographs, 200–750 mm Focal Length |
Omni-λ Series
| — | — | — | Contact |