OmniFluo900 Series Fluorescence Spectrometers, Steady-State and Lifetime
OmniFluo900 SeriesDesign Tools
Modular research fluorescence spectrometer for steady-state spectra and lifetimes from picoseconds to seconds.
The OmniFluo900 series is a modular fluorescence spectrometer for research in materials science, physics, chemistry, life sciences and agricultural science. It measures both steady-state spectra and fluorescence lifetimes, and can be configured for spectral coverage from the ultraviolet to the mid-infrared and for lifetimes from hundreds of picoseconds to seconds. Microscope and cryostat options add spatially resolved and temperature-dependent measurements.
Key Features
- Modular platform combining steady-state spectra and lifetime measurement from hundreds of picoseconds to seconds
- Water Raman SNR above 10,000:1 with single-photon counting detection
- Spectral coverage from UV to mid-IR (200–5500 nm) with optional PMT, InGaAs and InSb detectors
- 320 mm image-corrected monochromators with triple grating turrets; optional double monochromators for 1 × 10⁻⁹ stray light
- Spectral correction with factory-measured correction files as a standard procedure
- Large sample chamber with side laser port and options for polarizers, cryostats, integrating sphere and microscope
- Software for excitation, emission, synchronous and 3D scans plus kinetic, lifetime and TRES measurements with lifetime fitting
Typical Applications
- Semiconductor photoluminescence (PL) and PLE measurements
- Phosphor reflectance, absorption and quantum yield
- Upconversion fluorescence
- OLED and display materials
- Carbon nanotubes
- Chlorophyll and carotenoid analysis, biomedical fluorescence diagnostics
Related Products
- Finder 930 — Automated Confocal Raman Microscope
- RTS-mini — Confocal Raman System for Upright Microscopes
- Finder Insight Pro — Benchtop Raman Microscope
Browse all: Raman & Fluorescence Spectrometers · Spectroscopic Instruments
Models: OmniFluo960, OmniFluo990 · Contact WaveQuanta for a quote, configuration advice and lead time.
| Measurement Modes | Steady-state and time-resolved (lifetime) fluorescence |
| Spectral Range (Standard) | Emission 200–870 nm; excitation 200–800 nm |
| Extended Spectral Coverage | Up to 200–5500 nm with optional detectors |
| Spectral Resolution | 0.1 nm (system); 0.08 nm @ 435.84 nm with 1200 g/mm grating |
| Water Raman SNR | >10,000:1 (SQRT method) |
| Lifetime Range | 500 ps–10 s (TCSPC 500 ps–2.2 s; MCS 10 µs–10 s) |
| Monochromators | Czerny-Turner, 320 mm, image-corrected; optional 2 × 180 mm double monochromator |
| Stray Light | 1 × 10⁻⁵ (single); 1 × 10⁻⁹ (double monochromator) |
| Wavelength Accuracy / Repeatability | ±0.2 nm / ±0.1 nm |
| Minimum Wavelength Step | 0.005 nm |
| Grating Turret | Triple grating turret in each monochromator |
| Excitation Gratings | 1200 g/mm (300 nm blaze); 600 g/mm (500 nm blaze) |
| Emission Gratings | 1200 g/mm (500 nm blaze); 600 g/mm (750 nm blaze); 300 g/mm (1250 nm blaze) |
| Slits | Motorized, 0.01–3 mm continuously adjustable |
| Filter Wheels | 6-position at excitation and emission exits; automatic order-sorting filters |
| Steady-State Source | 75 W xenon arc lamp, 200–1800 nm, lamp-cup design, coupling efficiency over 90% |
| Pulsed Sources (Optional) | µs pulsed xenon flashlamp, ps pulsed laser or LED, supercontinuum, OPO tunable ns laser |
| Other Optional Sources | CW lasers (e.g. 808 nm, 980 nm), X-ray tube, lamp-pumped ns DPSS laser |
| Standard Detector | TE-cooled photon-counting PMT (−10 °C), 185–900 nm |
| Optional PMTs (UV–NIR) | TE-PMT-R13456: 185–980 nm; TE-PMT-R2658: 185–1010 nm |
| Optional PMTs (NIR) | TE-PMT-H10330C-75: 950–1700 nm; LN-PMT-R5509-73: 300–1700 nm |
| Optional Analog Detectors | TE-InGaAs 800–1700 nm or 800–2600 nm; LN-InSb 1000–5500 nm |
| Photon Counter | 100 Mcps, 1 MB/s sampling, 4 analog inputs (1–10 V), 16-bit A/D |
| TCSPC Module | 100 Mcps; 16–1024 ps resolution; 65,535 channels |
| Lock-in Amplifier | 50 mHz–120 kHz, dynamic reserve over 100 dB |
| Sample Chamber | Fused silica lenses; 50 × 50 or 25 × 25 mm filter holder; side laser port with variable ND |
| Sample Holders | Standard: cuvette, powder, film; optional rotating, stirred, water-bath heated |
| Cryostat (Optional) | 77–500 K or 3–300 K |
| Spectral Scan Modes | Excitation, emission, synchronous, 3D; optional anisotropy and temperature scans |
| Lifetime Modes and Analysis | Kinetic, lifetime and TRES scans; lifetime fitting; quantum yield; spectral correction |
Product Family: Raman & Fluorescence Spectrometers (7 variants)
| Product | SKU | Size | Clear Aperture | Operating Wavelength | |
|---|---|---|---|---|---|
| OmniFluo900 Series Fluorescence Spectrometers, Steady-State and Lifetime Current |
OmniFluo900 Series
| — | — | — | |
| Finder Edge Handheld Raman Spectrometer, 785 nm, 200–2000 cm⁻¹, 0.48 kg |
Finder Edge
| — | — | — | Contact |
| FI-FO Series Portable Raman Analyzers with Fiber Probe, 785/830/1064 nm |
FI-FO Series
| — | — | — | Contact |
| RTS-mini Confocal Raman System for Upright Microscopes, with Raman Mapping |
RTS-mini
| — | — | — | Contact |
| FI-Lite Portable Raman Analyzer for Liquids, 785 nm or 830 nm Excitation |
FI-Lite
| — | — | — | Contact |
| Finder Insight Pro Benchtop Raman Microscope, Built-in Sample Viewing |
Finder Insight Pro
| — | — | — | Contact |
| Finder 930 Automated Confocal Raman Microscope, 0.5 µm Lateral Resolution |
Finder 930
| — | — | — | Contact |