FL 26BW Long-Exposure Cooled Mono CMOS Camera, 1.8-inch, 6240 × 4168
FL 26BWDesign Tools
Deep-cooled monochrome camera for exposures of up to 60 minutes in luminescence and fluorescence imaging.
The FL 26BW is an air-cooled monochrome camera built around the Sony IMX571BLR-J back-illuminated CMOS sensor and designed for long-exposure work. The sensor is held at -25 °C (at 22 °C ambient) and dark current stays below 0.0005 e⁻/pixel/s, so clean, even images can be recorded with exposures of up to 60 minutes. Its 28.3 mm (1.8-inch) diagonal, 92% peak QE and sealed cooling chamber make it a practical alternative to cooled CCD cameras in chemiluminescence and similar low-signal applications.
Key Features
- Sony IMX571BLR-J BSI CMOS sensor: 92% peak QE at 530 nm and readout noise down to 0.85 e⁻ (Gain 3), with no amplifier glow
- Dark current below 0.0005 e⁻/pixel/s with the sensor cooled to -25 °C at 22 °C ambient; the sealed chamber prevents condensation
- Exposure times from 34 µs to 60 min for chemiluminescence and other low-signal imaging
- Four gain modes (full well from 49 ke⁻ to 3 ke⁻) and a 16000:1 dynamic range
- Defect pixel correction gives a uniform background for quantitative analysis
- Flexible binning from 2 × 2 up to 16 × 16 to raise sensitivity and dynamic range
- Compact 85 × 85 × 97 mm housing for instrument integration; supports LabVIEW, MATLAB and Micro-Manager 2.0
Typical Applications
- Chemiluminescence imaging
- Bioluminescence imaging
- Digital PCR (dPCR)
- Fluorescence imaging
- Biochip readout
Related Products
- FL 20BW — Mono Cooled CMOS Camera, 20 MP, 84% QE
- FL 20 — Cooled Color CMOS Camera, 20 MP, 1-inch Sensor, USB 3.0
- FL 9BW — Long-Exposure Cooled Mono CMOS Camera, 9 MP, 3000 × 3000
Browse all: Scientific Research Camera · Cameras
Model: FL 26BW · Acquisition software and drivers are supplied with the camera. Contact WaveQuanta for a quote and lead time.
| Sensor Type | BSI CMOS |
| Sensor | Sony IMX571BLR-J |
| Color / Mono | Mono |
| Peak QE | 92% @ 530 nm |
| Resolution | 6240 × 4168 |
| Pixel Size | 3.76 × 3.76 µm |
| Effective Area | 23.4 × 15.6 mm |
| Array Diagonal | 28.3 mm (1.8") |
| Full-Well Capacity | 49 / 15.7 / 7.8 / 3 ke⁻ (Gain 0 / 1 / 2 / 3) |
| Readout Noise | 2.7 / 1.0 / 0.95 / 0.85 e⁻ (Gain 0 / 1 / 2 / 3) |
| Dynamic Range | 16000:1 |
| Frame Rate | 6.5 / 3.4 fps (Standard / Low Noise) @ 6240 × 4168; 35.5 fps @ 3120 × 2084 (sensor bin) |
| Shutter Type | Rolling, Global Reset |
| Exposure Time | 34 µs – 60 min |
| Binning | 2 × 2, 3 × 3, 4 × 4, 5 × 5, 6 × 6, 8 × 8, 16 × 16 |
| ROI | Supported |
| Cooling Method | Air cooling |
| Cooling Temperature | -25 °C @ 22 °C ambient |
| Dark Current | < 0.0005 e⁻/pixel/s |
| Image Correction | Defect pixel correction (DPC) |
| Bit Depth | 16 bit |
| Data Interface | USB 3.0 |
| Trigger | Hardware, software; outputs: Exposure Start, Global, Readout End, High, Low; Hirose 6-pin |
| Optical Interface | M42 / customizable |
| Power Supply | 12 V / 8 A |
| Power Consumption | ≤ 50 W |
| Dimensions | 85 × 85 × 97 mm (H × W × L) |
| Weight | 945 g |
| Operating Environment | Working: 0–40 °C, 10–85% RH; storage: -10–60 °C, 0–85% RH |
| Operating System | Windows, Linux |
| Software / SDK Support | Acquisition and analysis software, LabVIEW, MATLAB, Micro-Manager 2.0 |